Customization: | Available |
---|---|
Humidity Display Range: | 0%Rh~99%Rh |
7g Display: | 8" Big LCD Display Screen |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Power supply |
AC220V-230V/50Hz or AC110V-120V/50-60Hz |
Max. Power output | 1300VA |
Mode of control | Air mode and baby mode controlled by micro-computer |
Air temperature control range | 25ºC-37ºC (over mode 37-39ºC) |
Baby temperature control range | 34ºC-37ºC (over mode 37-38ºC) |
Accuracy of skin temperature sensor | +-0.2ºC |
The temperature difference between incubator temperature and average incubator temperature (under steady temperature condition) | <=0.5ºC |
Temperature uniformity | <=0.8ºC (mattress is on the horizontal position) |
Inclination of the bassinet | +-12 |
Noise inside hood | <=45dB(A) |
Failure alarm | over-temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm, sensor box position alarm, tank position alarm, system alarm and so on |
Humidity display range | 0%RH~99%RH |
Humidity control range | 0%RH~90%RH |
Accuracy of humidity | +-10%RH |
Indication range of oxygen concentration | 0%~99% |
Control range of oxygen concentration | 20%~60% |
Control Accuracy of oxygen concentration | +-4% |
Range of weight | 100-8000g |
Resolution for weight indication | 1g |
Accuracy of weight | +-1% |
Operating temperature range | 20ºC~30ºC |
Ambient air movement rate | <0.3m/s |
Standard Configuration | Main body (including baby compartment, base, controller, I.V. POLE AND TRAY), the sensor box, skin temperature sensor and vertical height adjustment stand(VHA stand) |
Optional Configuration | Weighing system and monitor shelf, Disposable skin temperature sensor |