Customization: | Available |
---|---|
Type: | Ventilator |
Ventilation Mode: | Ippv, Sippv, Simv, Peep, Manual, Sigh (Imv, Vcv |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Display | 10.4" LCD display |
Ventilation mode | IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH (IMV, VCV, PCV) |
Tidal volume | 20~2000ml |
Rate | 1~99bpm |
SIMV rate | 1~20bpm |
I:E | 2:1~1:8 |
Inspiratory trigger pressure | -1.0~2.0 kPa |
PEEP | 0~2.0 kPa |
Pressure range | 1.0~6.0 kPa |
SIGH inspiratory time | 1.5 times |
Parameters for ventilation monitoring | Ventilation volume, tidal volume, SIMV rate, IPPV rate, total respiratory rate, peak pressure of airway, I/E, flow rate - time waveform, pressure - time waveform, inspiratory trigger pressure ,PEEP, Inspiratory platform |
Oxygen Concentration | 21%-100% |
Oxygen Concentration alarm | upper limit setting range 21%~100% low limit setting range 10%~80% |
Airway pressure alarm | upper limit setting range 0.3~6.0 kPa |
low limit setting range | 0.2~5.0 kPa |
Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min |
low limit setting range | 1.0~10L/min |
Sustained high-pressure alarm | > 2.5 kPa |
Suffocation alarm | no tidal volume input for 15 seconds |
Power alarm | Yes |
Power | AC /DC |
CO2, ECG monitoring function | optional |