Basic Info.

Model NO.
WHY7000D
Certification
CE, FDA, ISO13485
Group
for Lab and Clinic
Interface
1RS232 Serial Port
Filter
8 Filters
Printer
Thermal Paper, 40columns with Graphic Performances
Display
LCD Screen
Mains Supply
220V+-22V/50Hz±1 Hz 100va;Power <100W.
System
Open Reagent System
Storage
1000 Test Results
Transport Package
1 PCS/Carton
Specification
ISO/CE
Trademark
Poweam
Origin
China
HS Code
90189090
Production Capacity
5000PCS/Year

Product Description

LCD Display
8-wavelength
Reasonable software, programmable biochemical test menu, 200 test programmes can be saved
1000 patients test data can be saved
Clock CMOS chip inserted
Hermetic solid-state optical system, low contamination of dust and dampness,long life time of hard
membrane filters
Back spectro photometry, 8 hard membrane filter, separate signal process circuit for each wavelength,
filters select completed electronically,no mechanical movement
Titanium alloy flow cell with small capacity and minimize cross-contamination
Back-light LCD monitor and display character and curve simultaneously
Environment:Temperature :0°C-40°C
Self-test program, test error<1%, CV<3%
Humidity: 5-80%
Analysis method: kinetics, end point, two point, etc.
Power supply: 20w/12V halogen light, life time: 2000hr
Optical system: 340-800nm, 8 hard membrane filter: 340, 380, 405, 492, 510, 546, 578,
630nm(wavelength)
Accuracy: ±2nm
Precision: error<1%, drift<5mAbs/hr
Abs range: -2.500-2.5Abs, resolution:0.001Abs
Flow cell: titanium quartz alloy, 33uL; optical path: 10mm
Temperature: PELTIER elements, 25,30,37±0.1°C
200 test programmes can be saved
58mm thermal printer, no noise, high speed, easy maintenance
Keyboard: 24 keys touch keyboard
Power supply: 220V±15%, 50Hz, 70VA
Interface: standard RS232 serial interface
 
LIGHT RESOURCE 12V/20W halogen lamp, 2000hours life
OPTICS 8-wavelength static photometric group with 340,380,405,492,510,546,578,630nm interference filters
DETECTORS 8silicon photodiodes for UV/VIS
CUVETTE 33μl flow cell, temperature controlled by PELTER
THERMOSTAT Programmable for indoor, 25,30,37 ºC (±0.1ºC)
READING VOLUME 200-2000μl
ASPIRATION Precision pump with programmable volume
PRECISION/ACCURACY Precision1% at 2 O.D., photometric accuracy±1mABS,drift less than 5 m ABS/hour
KEYBOARD 24keys, membrane type
DISPLAY LCD type, graphic/alphanumeric 320×240 characters,
PRINTER thermal paper, 40columns with graphic performances
INTERFACE 1Rs232 serial port
MAINS SUPPLY 220V±22V/50Hz±1 Hz 100VA;Power <100W.
FUSE 1A, 2 pc
Clinic Biochemistry Analyzer Semi-Automatic Biochemistry Analyzer (WHY7000D)
Clinic Biochemistry Analyzer Semi-Automatic Biochemistry Analyzer (WHY7000D)
Clinic Biochemistry Analyzer Semi-Automatic Biochemistry Analyzer (WHY7000D)

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier