Below is EL-1200 Parameter:
Models:EL-1200A: K+, Na+
EL-1200B: K+, Na+, Cl-, HCO3, AG
EL-1200C: K+, Na+, Cl-, Ca2+, PH
EL-1200D: K+, Na+, Cl-, Ca2+, PH, HCO3, AG
EL-1200E: K+, Na+, Cl-, Li+
EL-1200F: K+, Na+, Cl-
Excellent precision and reliability
Long life, high performance, maintenance -free electrodes
Automatic sampling, washing and calibration
Supporting barcode reader
Auto sampler optional
Touch screen ,large LCD with backlight
Internal thermal recorder
Principle |
Direct measurement by Ion Selective Electrode (ISE) |
Sample |
Whole blood, serum, plasma, diluted urine |
Sample volume |
100~150µL |
Throughput |
up to 60 samples/h |
Data storage |
up to 200 patient result |
Calibration |
Automatic or On-demand |
Input |
Touch-screen |
Display |
Large LCD with backlight |
Output |
Internal thermal recorder, RS-232 serial port |
Temperature |
15ºC~32ºC, Relative humidity |
Input voltage |
AC 220V/110V ±10%, 50/60Hz |
Power consumption |
60W |
Option |
Auto sampler |
Items |
Measuring range |
Reproducibility(CV%) |
K+ |
0.50~15.00 mmol/L |
≤1.5% |
Na+ |
30.0~200.0 mmol/L |
≤1.0% |
Cl- |
30.0~200.0 mmol/L |
≤1.5% |
Ca2+ |
0.10~5.00 mmol/L |
≤2.0% |
Li+ |
0.20~3.00 mmol/L |
≤3.0% |
pH |
4.00~9.00 unit |
≤1.0% |